1.
Nucleus (La Habana) ; (46): 21-26, jul.-dic. 2009. ilus, graf, tab
Article in English | LILACS | ID: lil-738923

ABSTRACT

ABSTRACT The capabilities of the LowBackground Gamma Spectrometer (LBGS) at InSTEC were studied for environmental purposes. Fifty three glines were identified in the LBGS background spectrum. The Minimum Detectable Activity for , , , , and were calculated using the detector’s volumetric efficiency simulated by Monte Carlo method. Validation was performed by absolute and relative analysis of radionuclide activities present in a marine sediment certified material.


RESUMEN Se determinan las potencialidades del Espectrómetro Gamma de Bajo Fondo del InSTEC con fines ambientales. Se identificaron 53 líneas gamma en el espectro de fondo natural del espectrómetro. Se calculan las actividades mínimas detectables para los radionucleidos , , , , and empleando la eficiencia volumétrica del detector simulada por Monte Carlo. Como validación se determinan, por vía absoluta y relativa, las actividades de los radionucleidos presentes en un estándar de sedimento marino.